Thermoluminescent dosimeter


A thermoluminescent dosimeter, or TLD, is a type of radiation dosimeter, consisting of a piece of a thermoluminescent crystalline material inside a radiolucent package.
When a thermoluminescent crystal is exposed to ionizing radiation, it absorbs and traps some of the energy of the radiation in its crystal lattice. When heated, the crystal releases the trapped energy in the form of visible light, the intensity of which is proportional to the intensity of the ionizing radiation the crystal was exposed to. A specialized detector measures the intensity of the emitted light, and this measurement is used to calculate the approximate dose of ionizing radiation the crystal was exposed to.
Materials exhibiting thermoluminescence in response to ionizing radiation include calcium fluoride, lithium fluoride, calcium sulfate, lithium borate, calcium borate, potassium bromide, and feldspar. It was invented in 1954 by Professor Farrington Daniels of the University of Wisconsin-Madison.

Types

The two most common types of TLDs are calcium fluoride and lithium fluoride, with one or more impurities to produce trap states for energetic electrons. The former is used to record gamma exposure, the latter for gamma and neutron exposure. Other types include beryllium oxide, and calcium sulfate doped with thulium.
As the radiation interacts with the crystal it causes electrons in the crystal's atoms to jump to higher energy states, where they stay trapped due to intentionally introduced impurities in the crystal, until heated. Heating the crystal causes the electrons to drop back to their ground state, releasing a photon of energy equal to the energy difference between the trap state and the ground state.