Potentiostat


A potentiostat is the electronic hardware required to control a three electrode cell and run most electroanalytical experiments. A Bipotentiostat and polypotentiostat are potentiostats capable of controlling two working electrodes and more than two working electrodes, respectively.
The system functions by maintaining the potential of the working electrode at a constant level with respect to the reference electrode by adjusting the current at an auxiliary electrode. It consists of an electric circuit which is usually described in terms of simple op amps.

Primary use

This equipment is fundamental to modern electrochemical studies using three electrode systems for investigations of reaction mechanisms related to redox chemistry and other chemical phenomena. The dimensions of the resulting data depend on the experiment. In voltammetry, electric current in amps is plotted against electric potential in voltage. In a bulk electrolysis total coulombs passed is plotted against time in seconds even though the experiment measures electric current over time. This is done to show that the experiment is approaching an expected number of coulombs.
Most early potentiostats could function independently, providing data output through a physical data trace. Modern potentiostats are designed to interface with a personal computer and operate through a dedicated software package. The automated software allows the user rapidly to shift between experiments and experimental conditions. The computer allows data to be stored and analyzed more effectively, rapidly, and accurately than historic methods.

Basic relationships

A potentiostat is a control and measuring device. It comprises an electric circuit which controls the potential across the cell by sensing changes in its resistance, varying accordingly the current supplied to the system: a higher resistance will result in a decreased current, while a lower resistance will result in an increased current, in order to keep the voltage constant as described by Ohm's law.
As a result, the variable system resistance and the controlled current are inversely proportional

Principles of operation

Since 1942, when Hickling built the first three electrode potentiostat, substantial progress has been made to improve the instrument. Hickling's device used a third electrode, the reference electrode to control the cell potential automatically. Up until the present day his principle has remained in use. At a glance, a potentiostat measures the potential difference between the working and the reference electrode, applies the current through the counter electrode and measures the current as an voltage drop over a series resistor.
The control amplifier is responsible for maintaining the voltage between the reference and the working electrode as closely as possible to the voltage of the input source. It adjusts its output to automatically control the cell current so that a condition of equilibrium is satisfied. The theory of operation is best understood using the equations below.
Prior to observing the following equations, one may note that, from an electrical point of view, the electrochemical cell and the current measurement resistor may be regarded as two impedances. includes in series with the interfacial impedance of the counter electrode and the solution resistance between the counter and the reference.
represents the interfacial impedance of the working electrode in series with the solution resistance between the working and the reference
electrodes.
The role of the control amplifier is to amplify the potential difference between the positive input and the negative input. This may be translated mathematically into the following equation:
where is the amplification factor of the CA. At this point the assumption may be made that a negligible amount of current is flowing through the reference electrode. This correlates to physical phenomenon since the reference electrode is connected to a high impedance electrometer. Thus, the cell current may be described in two ways:
and
Combining Eqs. and yields Eq. :
where is the fraction of the output voltage of the control amplifier returned to its negative input; namely the feedback factor:
Combining Eqs. and yields Eq. :
When the quantity becomes very large with respect to one, Eq. reduces to Eq., which is one of the negative feedback equations:
Eq. proves that the control amplifier works to keep the voltage between the reference and the working close to the input source voltage.

Software control

Replacing the CA, a control algorithm can maintain a constant voltage between the reference electrode and the working electrode. This algorithm is based on the rule of proportion:
If the measurement intervals of Eq. are kept constant, the control algorithm sets the cell voltage so to keep as close as possible to the setpoint. The algorithm requires software-controllable hardware such as a digital multimeter, a power supply, and a double-pole double-throw relay. The relay is necessary to switch polarity.

Significant features

In electrochemical experiments the electrodes are the pieces of equipment that comes in immediate contact with the analyte. For this reason the electrodes are very important for determining the experimental result. The electrode surface may or may not catalyze chemical reactions. The size of the electrodes affects the magnitude of the currents passed which can affect signal to noise. But electrodes are not the only limiting factor for electrochemical experiments, the potentiostat also has a limited range of operation. The following are a few significant features that vary between instruments.