Medipix


Medipix is a family of photon counting and particle tracking pixel detectors developed by an international collaboration, hosted by CERN.

Design

These are hybrid detectors as a semi-conductor sensor layer is bonded on to a processing electronics layer.
The sensor layer is a semiconductor, such as silicon, GaAs, or CdTe in which the incident radiation makes an electron/hole cloud. The charge is then collected to pixel electrodes and via bump bonds conducted to the CMOS electronics layer.
The pixel electronics first amplifies the signal and then compares the signal amplitude with a pre-set discrimination level. The subsequent signal processing depends on type of the device. Standard Medipix detector increases the counter in the appropriate pixel if the signal is above the discrimination level. The Medipix devices contain also an upper discrimination level and hence only signals within a range of amplitude could be accepted.
The Timepix devices offer in addition to the counting two more modes of operation. The first one is so called “Time-over-Threshold” mode. It is a mode where the counter in each pixel records number of clocks for which the pulse remains above the discrimination level. This number is proportional to the energy of detected radiation. This mode is useful for particle tracking applications or for direct spectral imaging.
The second mode of the Timepix chip is “Time-of-arrival” in which pixel counters record time between a trigger and detection of radiation quanta with energy above the discrimination level. This mode of operation finds use in Time of flight applications for instance in neutron imaging.
Every individual hit of radiation is processed by the electronics integrated in each pixel this way, therefore the device could be considered as 65 536 individual counting detectors or even spectrometers. The energy discriminators are adjustable. Therefore, scanning with their level is possible to measure the full spectrum of the incoming radiation and thus enabling full spectroscopic x-ray imaging.
Medipix-2, Timepix, and Medipix-3 are all 256×256 pixels, each 0.055mm square, forming a total area 14.08mm × 14.08mm. Larger area detectors are possible to crate by bump-bonding multiple chips to larger monolithic sensors. Detectors of sizes from 2x2 to 2x4 chips are commonly used. Even larger gap-less areas could be created using the technology of edgeless sensors. Here the Medipix/Timepix chips have each its own sensor and these assemblies are tiled next to each other to create nearly arbitrarily sized detector array.

Comparison with existing technologies

Photon counting pixel detectors represent the next generation of radiation imaging detectors. The photon counting technology overcomes limitations of current imaging devices. Comparison of photon counting with existing technologies is in the following table:
Film emulsionsCharge integrating devicesPhoton counting pixel detectors
Operation principleChange of chemical or physical properties after interaction with radiation. Needs special treatment.Ionizing radiation creates light and subsequently an electric charge that is collected and integrated in pixels.Ionizing radiation creates charge directly in the sensor. The charge is compared with threshold and counted digitally in pixels.
AdvantagesVery high resolution, low noise, cheap.High spatial resolution. Low price.Good spatial resolution, high read-out speed, no noise, no dark current, unlimited dynamic scale, energy discrimination
DisadvantagesNonlinear response, limited dynamic scale, needs processingDark current, noise, limited dynamic scaleHigh price.

Versions

Medipix-1 was the first device of the Medipix family. It had 64x64 pixels of 170 µm pitch. Pixels contained one comparator with 3-bit per-pixel offset adjustment. The minimum threshold was ~5.5 keV. The counter depth was 15-bit. The maximum count rate per pixel was 2 MHz per pixel.
Medipix-2 is the successor of Medipix-1. The pixel pitch was reduced to 55 µm and the pixel array is of 256x256 pixels. Each pixel has two discrimination levels each adjustable individually in pixels using a 3-bit offset. The maximum count rate is about 100 kHz per pixel.
Medipix-2 MXR is an improved version of Medipix-2 device with better temperature stability, pixel counter overflow protection, increased radiation hardness and many other improvements.
Timepix is device conceptually originating from Medipix-2. It adds two more modes to the pixels, in addition to counting of detected signals: Time-over-Threshold and Time-of-Arrival. The detected pulse height is recorded in the pixel counter in the TOT mode. The TOA mode measures time between trigger and arrival of the radiation into each pixel.
Medipix-3 is the latest generation of photon counting devices for X-ray imaging. The pixel pitch remains the same as well as the pixel array size. It has better energy resolution through real time correction of charge sharing. It also has multiple counters per pixel that can be used in several different modes. This allows for continuous readout and up to eight energy thresholds.
Timepix-3 is a successor of the Timepix chip. One of the biggest distinguishing changes is the approach to the data readout. All previous chips utilized the frame-based readout, i.e. the whole pixel matrix was read out at once. Timepix-3 has event-based readout where values recorded in pixels are read out immediately after the hit together with coordinates of the hit pixel. The chip therefore generates a continuous stream of data rather than a sequence of frames. The next major difference compared to the previous Timepix chip is the ability to measure the hit amplitude simultaneously with the time of arrival. Other parameters such as energy and timing resolution were also improved compared to the original Timepix chip.

Readout electronics

The digital data recorded by Medipix/Timepix devices are transferred to a computer via readout electronics. The readout electronics is also responsible for setup and control of the detector parameters. Several readout systems were developed within the Medipix collaboration

Muros

Muros was one of the first readout systems of Medipix detectors. Muros was developed at Nikhef, Amsterdam, The Netherlands. It was relatively compact readout enabling access to all features of the detector. It allowed maximum frame rate of cca 30 frames/s with a single chip.

USB interface

This electronics was developed at IEAP-CTU, Czech Republic. It provides a lower frame rate compared to Muros, but the electronics was integrated into a box not larger than a pack of cigarettes. Moreover, no special PC hardware card was needed as it was in case of Muros. Therefore, the USB interface become quickly the most used readout within the Medipix collaboration and its partners.

Relaxd

Relaxd is a readout electronics developed at Nikhef. The data is transferred to PC via 1Gbit/s Ethernet connection. The maximum frame rate is at level of 100 frame/s.

Fitpix

Fitpix is the next generation of the USB interface developed by the group in Prague. The electronics implements the parallel Medipix/Timepix readout and therefore the maximum frame rate reaches 850 frame/s. It supports also the serial readout with frame rate of 100 frames/s.

Minipix

Minipix is a miniaturized integrated chip+readout electronics device developed by in Prague. The whole system has size of a USB flash drive. Several of these devices were used in International Space Station as radiation monitoring systems.

Spidr

Spidr is powerful readout electronics for the Timepix-3 chip. It is under development at Nikhef.

Excalibur and Merlin systems

Both systems are developed at Diamond Light Source, UK, for Medipix3 readout and applications at synchrotrons. Merlin is available with CdTe sensors from Quantum Detectors who are collaborating on further development with Diamond Light Source.

LAMBDA system

Lambda is a high-speed big area readout systems developed at DESY. Lambda is available with high-Z sensor options, such as GaAs and CdTe.

MARS

MARS is a gigabit Ethernet readout accommodating up to 6 Medipix 2 or Medipix 3 detectors. The electronics was developed at University of Otago, Christchurch, New Zealand.

Applications

X-ray imaging

is the primary application field of Medipix detectors. Medipix offers to the X-ray imaging field in particular an advantage in higher dynamic range and energy sensitivity. Examples of X-ray images from selected X-ray imaging application fields are:

Space radiation dosimetry

Timepix-based detectors from the Medipix2 Collaboration have been flown on the International Space Station since 2013, and on the first flight test of NASA's new Orion Multi-Purpose Crew Vehicle in December 2014. Current plans call for similar devices to be flown as the primary radiation area monitors on the future initial manned Orion missions.

Other

The detectors may also find applications in astronomy, high energy physics, medical imaging, and X-ray spectroscopy.

History