Hemocytometer


The hemocytometer is a counting-chamber device originally designed and usually used for counting blood cells.
The hemocytometer was invented by Louis-Charles Malassez and consists of a thick glass microscope slide with a rectangular indentation that creates a chamber. This chamber is engraved with a laser-etched grid of perpendicular lines. The device is carefully crafted so that the area bounded by the lines is known, and the depth of the chamber is also known. By observing a defined area of the grid, it is therefore possible to count the number of cells or particles in a specific volume of fluid, and thereby calculate the concentration of cells in the fluid overall. A well used type of hemocytometer is the Neubauer counting chamber.

Principles

The gridded area of the hemocytometer consists of nine 1 x 1 mm squares. These are subdivided in three directions; 0.25 x 0.25 mm, 0.25 x 0.20 mm and 0.20 x 0.20 mm. The central square is further subdivided into 0.05 x 0.05 mm squares. The raised edges of the hemocytometer hold the coverslip 0.1 mm off the marked grid, giving each square a defined volume.
DimensionsAreaVolume at 0.1 mm depth
1 x 1 mm1 mm2100 nL
0.25 x 0.25 mm0.0625 mm26.25 nL
0.25 x 0.20 mm0.05 mm25 nL
0.20 x 0.20 mm0.04 mm24 nL
0.05 x 0.05 mm0.0025 mm20.25 nL

Usage

To use the hemocytometer, first make sure that the special coverslip provided with the counting chamber is properly positioned on the surface of the counting chamber. When the two glass surfaces are in proper contact Newton's rings can be observed. If so, the cell suspension is applied to the edge of the coverslip to be sucked into the void by capillary action which completely fills the chamber with the sample. The number of cells in the chamber can be determined by direct counting using a microscope, and visually distinguishable cells can be differentially counted. The number of cells in the chamber is used to calculate the concentration or density of the cells in the mixture the sample comes from. It is the number of cells in the chamber divided by the chamber's volume, which is known from the start, taking account of any dilutions and counting shortcuts:
where the volume of the diluted sample divided by the volume of the original mixture in the sample is the dilution factor. For example, if the volume of the original mixture was 20μL and it was diluted once, then the second term in parentheses is 40μL/20μL. The volume of the squares counted is the one shown in the table at the top, depending on the size. The number of cells counted is the sum of all cells counted across squares in one chamber. The proportion of the cells counted applies if not all inner squares within a set square are counted.
For most applications, the four large corner squares are only used. The cells that are on or touching the top and left lines are counted, but the ones on or touching the right or bottom lines are ignored.

Requirements

The original suspension must be mixed thoroughly before taking a sample. This ensures the sample is representative, and not just an artifact of the particular region of the original mixture it was drawn from.
An appropriate dilution of the mixture with regard to the number of cells to be counted should be used. If the sample is not diluted enough, the cells will be too crowded and difficult to count. If it is too dilute, the sample size will not be enough to make strong inferences about the concentration in the original mixture.
By performing a redundant test on a second chamber, the results can be compared. If they differ more than 2 times the counting error, the method of taking the sample may be unreliable.
The counting chamber should be filled by capillary action after the chamber lid has been put in place. This avoids the risk that the cells could sediment/stick to the glass or some volume to evaporate before the cover-slip is placed on top and resulting in an overestimation of the cell concentration. Sedimentation is less of a problem with bacteria but evaporation, more prevalent in low-humidity air-conditioned laboratories, still has to be minimized.

Applications