Cryostat


A cryostat is a device used to maintain low cryogenic temperatures of samples or devices mounted within the cryostat. Low temperatures may be maintained within a cryostat by using various refrigeration methods, most commonly using cryogenic fluid bath such as liquid helium. Hence it is usually assembled into a vessel, similar in construction to a vacuum flask or Dewar. Cryostats have numerous applications within science, engineering, and medicine.

Types

Closed-cycle cryostats

Closed-cycle cryostats consist of a chamber through which cold helium vapour is pumped. An external mechanical refrigerator extracts the warmer helium exhaust vapour, which is cooled and recycled. Closed-cycle cryostats consume a relatively large amount of electrical power, but need not be refilled with helium and can run continuously for an indefinite period. Objects may be cooled by attaching them to a metallic coldplate inside a vacuum chamber which is in thermal contact with the helium vapour chamber.

Continuous-flow cryostats

Continuous-flow cryostats are cooled by liquid cryogens from a storage dewar. As the cryogen boils within the cryostat, it is continuously replenished by a steady flow from the storage dewar. Temperature control of the sample within the cryostat is typically performed by controlling the flow rate of cryogen into the cryostat together with a heating wire attached to a PID temperature control loop. The length of time over which cooling may be maintained is dictated by the volume of cryogens available.
Owing to the scarcity of liquid helium, some laboratories have facilities to capture and recover helium as it escapes from the cryostat, although these facilities are also costly to operate.

Bath cryostats

Bath cryostats are similar in construction to vacuum flasks filled with liquid helium. A coldplate is placed in thermal contact with the liquid helium bath. The liquid helium may be replenished as it boils away, at intervals between a few hours and several months, depending on the volume and construction of the cryostat. The boil-off rate is minimised by shielding the bath with either cold helium vapour, or vacuum shield with walls constructed from so-called super insulator material. The helium vapour which boils away from the bath very effectively cools thermal shields around the outside of the bath. In the older designs there may be additional liquid nitrogen bath, or several concentric layers of shielding, with gradually increasing temperatures. However, the invention of super insulator materials has made this technology obsolete.

Multistage cryostats

In order to achieve temperature lower than liquid helium additional cooler stages may be added to the cryostat.
Temperatures down to 1K can be reached by attaching the coldplate to a 1-K pot, which is a container of He-3 isotope which is connected to vacuum pump.
Temperatures down to 1mK can be reached by employing dilution refrigerator or dry dilution refrigerator typically in addition to the main stage and 1K pot.
Temperatures below that can be reached using magnetic refrigeration.

Applications

Magnetic resonance imaging and research magnet types

Cryostats used in MRI machines are designed to hold a cryogen, typically helium, in a liquid state with minimal evaporation. The liquid helium bath is designed to keep the superconducting magnet's bobbin of superconductive wire in its superconductive state. In this state the wire has no electrical resistance and very large currents are maintained with a low power input. To maintain superconductivity, the bobbin must be kept below its transition temperature by being immersed in the liquid helium. If, for any reason, the wire becomes resistive, i.e. loses superconductivity, a condition known as a "quench", the liquid helium evaporates, instantly raising pressure within the vessel. A burst disk, usually made of carbon, is placed within the chimney or vent pipe so that during a pressure excursion, the gaseous helium can be safely vented out of the MRI suite. Modern MRI cryostats use a mechanical refrigerator to re-condense the helium gas and return it to the bath, to maintain cryogenic conditions and to conserve helium.
Typically cryostats are manufactured with two vessels, one inside the other. The outer vessel is evacuated with the vacuum acting as a thermal insulator. The inner vessel contains the cryogen and is supported within the outer vessel by structures made from low-conductivity materials. An intermediate shield between the outer and inner vessels intercepts the heat radiated from the outer vessel. This heat is removed by a cryocooler. Older helium cryostats used a liquid nitrogen vessel as this radiation shield and had the liquid helium in an inner, third, vessel. Nowadays few units using multiple cryogens are made with the trend being towards 'cryogen-free' cryostats in which all heat loads are removed by cryocoolers.

Biological microtome type

Cryostat are used in medicine to cut histological slides. They are usually used in a process called frozen section histology. The cryostat is essentially an ultrafine "deli-slicer", called a microtome, placed in a freezer. The cryostat is usually a stationary upright freezer, with an external wheel for rotating the microtome. The temperature can be varied, depending on the tissue being cut - usually from minus 20 to minus 30 degree Celsius. The freezer is either powered by electricity, or by a refrigerant like liquid nitrogen. Small portable cryostats are available and can run off generators or vehicle inverters. To minimize unnecessary warming all necessary mechanical movements of the microtome can be achieved by hand via a wheel mounted outside the chamber. Newer microtomes have electric push button advancement of the tissue. The precision of the cutting is in micrometres. Tissue are sectioned as thin as 1 micrometre. Usual histology slides are mounted with a thickness of about 7 micrometres. Specimens that are soft at room temperature are mounted on a cutting medium on a metal "chuck", and frozen to cutting temperature. Once frozen, the specimen on the chuck is mounted on the microtome. The crank is rotated and the specimen advances toward the cutting blade. Once the specimen is cut to a satisfactory quality, it is mounted on a warm clear glass slide, where it will instantaneously melt and adhere. The glass slide and specimen is dried with a dryer or air dried, and stained. The entire process from mounting to reading the slide takes from 10 to 20 minutes, allowing rapid diagnosis in the operating room, for the surgical excision of cancer. The cryostat can be used to cut histology and tissue slide outside of medicine, but the quality of the section is poor compared to standard fixed section wax mounted histology. Newer technology such as the Compresstome, a type of vibrating microtome, utilizes agarose tissue embedding instead of an optimal cutting temperature compound to eliminate the need for traditional cryostat freezing and may be used for improved quality sectioning.