Capillary electrophoresis–mass spectrometry is an analytical chemistry technique formed by the combination of the liquid separation process of capillary electrophoresis with mass spectrometry. CE-MS combines advantages of both CE and MS to provide high separation efficiency and molecular mass information in a single analysis. It has high resolving power and sensitivity, requires minimal volume and can analyze at high speed. Ions are typically formed by electrospray ionization, but they can also be formed by matrix-assisted laser desorption/ionization or other ionization techniques. It has applications in basic research in proteomics and quantitative analysis of biomolecules as well as in clinical medicine. Since its introduction in 1987, new developments and application has made CE-MS powerful separation and identification technique. Use of CE-MS has increased for protein and peptides analysis and other biomolecules. However, the development of online CE-MS is not without challenges. Understanding of CE, the interface setup, ionization technique and mass detection system is important to tackle problems while coupling capillary electrophoresis to mass spectrometry.
is a separation technique which uses high electric field to produce electroosmotic flow for separation of ions. Analytes migrate from one end of capillary to other based on their charge, viscosity and size. Higher the electric field, greater is the mobility. Mass spectrometry is an analytical technique that identifies chemical species depending on their mass-to-charge ratio. During the process, an ion source will convert molecules coming from CE to ions that can then be manipulated using electric and magnetic field. The separated ions are then measured using a detector. The major problem faced when coupling CE to MS arises due to insufficient understanding of fundamental processes when two techniques are interfaced. The separation and detection of analytes can be improved with better interface. CE has been coupled to MS using various ionization techniques like FAB, ESI, MALDI, APCI and DESI. The most used ionization technique is ESI.
The first not CE-MS interface had cathode end of CE capillary terminated within a stainless steel capillary. An electrical contact was made at that point completing the circuit and initiating the electrospray. This interface system had few drawbacks like mismatch in the flow rates of two systems. Since then, interface system has been improved to have continuous flow rate and good electrical contact. At present, three types of interface system exist for CE/ESI-MS which are discussed briefly.
Sheathless interface
CE capillary is coupled directly to an electrospray ionization source with a sheathless interface system. The electric contact for ESI is realized by using capillary coated with conductive metal. Because no sheath liquid is used, the system has high sensitivity, low flow rates and minimum background. However, these interface designs, all have challenges including low mechanical robustness, poor reproducibility. The latest sheathless interface design features porous ESI emitter through chemical etching. This design effectively provides robust interfacing with mass spectrometry and addresses the reproducibility challenges associated with previous designs. This porous emitter interface has been explored to couple of CITP/CZE which greatly improves sample loading capacity of CE and enabled ultrasensitive detection of trace analytes. Another key factor for successful CE-MS interface is the choice of buffer solution which must be suitable for both CE separation and ESI operation.
Sheath flow interface
With the sheath flow interface, the electrical connection is established when the CE separation liquid is mixed with sheath liquid flowing coaxially in a metal capillary tubing. Commonly used sheath liquid is 1:1 mixture of water-methanol with 0.1% acetic acid or formic acid. The system is more reliable and has wide selection range of separation electrolyte. There might be some decrease in sensitivity due to sheath liquid.
Liquid junction interface
This technique uses a stainless steel tee to mix separation electrolyte from CE capillary with make up liquid. The CE capillary and ESI needle are inserted through opposite sides of the tee and a narrow gap is maintained. The electrical contact is established by makeup liquid surrounding the junction between two capillaries. This system is easy to operate. However, the sensitivity is reduced and the mixing of two liquids could degrade separation.
CE can be coupled to fast atom bombardment ionization using a continuous flow interface. The interface must match the flow rate between the two systems. The CF-FAB requires a relatively high flow rate but CE need low flow rate for better separation. A make-up flow can be used using a sheath flow or liquid junction.
Coupling CE with MALDI-MS
Off-line coupling of CE to MALDI, the CE effluent could be sprayed or added drop wise on MALDI target plate then dried and analyzed by MS. For online coupling, a moving target with continuous contact to CE capillary end is required. The moving target takes analytes into MS where it is desorbed and ionized. Musyimi et al. developed a new technique where rotating ball was used to transfer CE to MS. The sample from CE is mixed with matrix coming though another capillary. As the ball rotates the sample is dried before it reaches ionization region. This technique has high sensitivity since no makeup fluid is used.
Applications
CE-MS ability to separate analytes present in extremely low concentration with high efficiency at high speed has made it applicable in all fields of science. CE-MS has been used for bioanalytical, pharmaceuticals, environmental and forensic application. The major application of CE-MS has been for biological studies, mostly for protein and peptide analysis. Along with that, it is used often for routine analysis of pharmaceutical drugs. There are number of studies reporting characterization of mixtures of peptides and proteins.CE-MS can be used for routine clinical checkup. Body fluids like blood and urine have been analyzed with CE-MS to identify biomarkers for renal diseases and cancer.